The MG15 is able to support four active gauges and up to three passive gauges, extending the measurement range to 2×10-12 mbar. The unit is fully software controlled. The unit can be remotely controlled via one of available interfaces.
  

Includes communication interface RS-232/485 and Ethernet IP 

Cables not included, depends on the sensors. 

A complete List with Prices and which Cables for which compatible Gauges are available are under Specials & Download’s. manufactured by PREVAC 

Product DESCRIPTION

MG15 ION MULTI GAUGE CONTROLLER

The MG15 is able to support four active gauges and up to three passive gauges, extending the measurement range to 2×10-12 mbar.

The unit is fully software controlled. The unit can be remotely controlled via one of available interfaces.

 

Features

• measurement range from atmospheric to 2x10-12 mbar

• pressure plots

• pressure trend graphs

• measurement filtering (low, med, hi)

• controls almost all commercially available Bayard-Alpert gauges

• degassing of Bayard-Alpert with power and time limit

• supports Ir/W/Thoria filaments

• Bayard-Alpert overpressure protection

• selectable measurement units: mbar, Torr, Pascal, Psia

• gas specific correction with one customized setting

• unit over-temperature protection

• 10 individually programmable set-points with threshold and hysteresis functions

• user-defined channel names

• possibility of active gauges self-identification

• charts module - real time pressure charts with data export possibility (option)

• pararell measurement for all channels

• flux monitor

• black box version available (option) - without 7" TFT display

• fully software controlled (option)

• autosave function for saving your parameters/unit pre-set and applying them automatically after restart.

• Labview compatible (option)

 

Applications

• Any high and ultra high vacuum systems

• Measurement of pressure (from atmosphere to UHV range)

• Bakeable and non bakeable systems

Product Details

Created with Sketch.

Technical data

Supply voltage              100 - 240 VAC, 50 - 60 Hz (power consumption max 200 W)

Measurement channels                  up to 7 (active channels: 4; passive channels: up to 3 - number defined at the time of order)

Supported active gauges                CTR90, TTR91, TTR211, PTR225, PKR251/360/361, PCR280, PTR90, ITR90, ITR100, Baratron, ANALOG IN, PG105, ATMION, User defined*

Supported passive gauges             IE414, IE514, UHV-24, UHV-24p, MKS Series 274, NUDE-UHV-8A, NR-F-UHV, G8130, BARION basic II, User defined*

Measurement range:

Passive channel

Active channel              

2x12-12 mbar

5x10-10 mbar

Display range               2x10-12 - 2x105mbar

Setpoints relay outputs                   10 (4xNO/NC, 6xNO)

Relay outputs with configurable activation signal                4 (Emission, Degas)

Digital inputs                 4

Analog outputs             4 (0-10V)

Communication interface                RS232/485, Ethernet, PROFINET (option - specify at order)

Communication protocol                 MODBUS-TCP

User interface               7" TFT display with touchscreen

Dimensions [mm]         212.6 x 128.4 x 260.3 mm (WxHxD), 1/2 of 3HE 19"

Weight (approx.)           3.8 kg

 

* prepared to support many other custom gauges